Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

FDA recall source

Recall number
Z-0002-2027
Event id
99843
Status
Ongoing
FDA classification
Class I
Recalling firm
Instrumentation Laboratory
Report date
Oct 21, 2026
Posted
Oct 10, 2026
Initiated
Sep 3, 2026
Center classification date
Oct 9, 2026
Product code
LCO
Device type
Platelet Factor 4 Radioimmunoassay
Reason
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Root cause
Under Investigation by firm
Quantity
1,414 units
Code info
Part Number: 0009802028. UDI-DI: 08426950545774. Lot Number (Expiration Date): B40006 (01/16/2027), B38063 (11/01/2026).
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K170854
Location
Bedford, MA, United States