Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

FDA recall source

Recall number
Z-0070-2027
Event id
99704
Status
Ongoing
FDA classification
Class II
Recalling firm
Steris Corporation
Report date
Oct 14, 2026
Posted
Oct 2, 2026
Initiated
Aug 31, 2026
Center classification date
Oct 2, 2026
Product code
FRC
Device type
Indicator, Biological Sterilization Process
Reason
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Root cause
Under Investigation by firm
Quantity
266
Code info
UDI: 10724995220324 Lot numbers: 260810 - expiration date 08/10/2026 260810A - expiration date 08/10/2026 261106 - expiration date 11/06/2026 270116 - expiration date 01/16/2027
Distribution
Worldwide - US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K854902
Location
Mentor, OH, United States