Device recall

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

FDA recall source

Recall number
Z-0073-2027
Event id
99704
Status
Ongoing
FDA classification
Class II
Recalling firm
Steris Corporation
Report date
Oct 14, 2026
Posted
Oct 2, 2026
Initiated
Aug 31, 2026
Center classification date
Oct 2, 2026
Product code
FRC
Device type
Indicator, Biological Sterilization Process
Reason
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Root cause
Under Investigation by firm
Quantity
2205
Code info
UDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027
Distribution
Worldwide - US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K083665
Location
Mentor, OH, United States