Device recall

DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520

FDA recall source

Recall number
Z-0084-2027
Event id
99800
Status
Ongoing
FDA classification
Class II
Recalling firm
Beckman Coulter Inc.
Report date
Oct 14, 2026
Posted
Oct 2, 2026
Initiated
Sep 2, 2026
Center classification date
Oct 2, 2026
Product code
JJE
Device type
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
Root cause
Software design
Quantity
146 US, 170 OUS
Code info
Software version 1.4.5 or earlier. 1. UDI-DI 14987666545058 2. UDI-DI 14987666545065
Distribution
Worldwide - US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K220977
Location
Brea, CA, United States