Device recall

Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No

FDA recall source

Recall number
Z-0089-2027
Event id
99549
Status
Ongoing
FDA classification
Class II
Recalling firm
Nihon Kohden America LLC
Report date
Oct 14, 2026
Posted
Oct 6, 2026
Initiated
Sep 8, 2026
Center classification date
Oct 6, 2026
Product code
MHX
Device type
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason
Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
Root cause
Software design
Quantity
808
Code info
Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K102376
Location
Irvine, CA, United States