Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;

FDA recall source

Recall number
Z-0107-2027
Event id
99810
Status
Ongoing
FDA classification
Class II
Recalling firm
POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC
Report date
Oct 14, 2026
Posted
Oct 7, 2026
Initiated
Aug 25, 2026
Center classification date
Oct 7, 2026
Product code
JPH
Device type
Test, Erythrocyte Sedimentation Rate
Reason
During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).
Root cause
Process control
Quantity
448400
Code info
Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27
Distribution
Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K840500
Location
Cortlandt Manor, NY, United States