Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

ePM Series Patient Monitor. Patient Monitor Series: ePM 10M/ePM 12M/ePM 12MA/ ePM 15M/ePM 10A

FDA recall source

Recall number
Z-0109-2027
Event id
99828
Status
Ongoing
FDA classification
Class III
Recalling firm
Mindray DS USA, Inc.
Report date
Oct 14, 2026
Posted
Oct 7, 2026
Initiated
Aug 31, 2026
Center classification date
Oct 7, 2026
Product code
MHX
Device type
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Root cause
Labeling design
Quantity
32,184 units
Code info
Part numbers of ePM Series Patient Monitors Operator's Manuals: 1. Guardian Plus Operator's Manual (OEM, MDPro) up to and including revision 4.0; Part No. 046-026408-00 2. ePM 10A Operator's Manual up to and including revision 12.0; Part No. 046-019796-00 3. ePM 12MA Operator's Manual up to and including revision 13.0; Part No. 046-019797-00 4. ePM 10M, ePM 12M Operator's Manual up to and including revision 13.0; Part No. 046-019798-00 5. ePM 15M Operator's Manual up to and including revision 3.0; Part No. 046-028915-00
Distribution
Worldwide - Nationwide and the country of Canada.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K260098
Location
Mahwah, NJ, United States