Device recall
Panorama and Telemetry Monitoring System. Operator's Manuals: TMS-6016/Panorama Telepack-608/TMS60 (TD60).
- Recall number
- Z-0112-2027
- Event id
- 99828
- Status
- Ongoing
- FDA classification
- Class III
- Recalling firm
- Mindray DS USA, Inc.
- Report date
- Oct 14, 2026
- Posted
- Oct 7, 2026
- Initiated
- Aug 31, 2026
- Center classification date
- Oct 7, 2026
- Product code
- MSX
- Device type
- System, Network And Communication, Physiological Monitors
- Reason
- Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
- Root cause
- Labeling design
- Quantity
- 11,831 units
- Code info
- Part Numbers of the Panorama and Telemetry Monitoring System Operator's Manuals: 1. Panorama Telepack-608 Telemetry Monitoring System Operator's Manual up to and including revision 1.0; Part No. 046-007526-00 2. TMS-6016 Operator's Manual up to and including revision 11.0; Part No. 046-005012-00
- Distribution
- Worldwide - Nationwide and the country of Canada.
- Voluntary or mandated
- Voluntary: Firm initiated
- Related 510(k)s
- K242728
- Location
- Mahwah, NJ, United States