Device recall
Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are malleable and can be molded or cut into various sizes and shapes according to the intended implant site.
- Recall number
- Z-0137-2027
- Event id
- 99701
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Report date
- Oct 21, 2026
- Posted
- Oct 9, 2026
- Initiated
- Aug 12, 2026
- Center classification date
- Oct 9, 2026
- Product code
- MBP
- Device type
- Filler, Bone Void, Osteoinduction (W/O Human Growth Factor)
- Reason
- Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection
- Root cause
- Process change control
- Quantity
- 18
- Code info
- UDI-DI: 00643169123021. Serial Numbers: A62881-034 through A62881-051, specifically: A62881-034, A62881-035, A62881-036, A62881-037, A62881-038, A62881-039, A62881-040, A62881-041, A62881-042, A62881-043, A62881-044, A62881-045, A62881-046, A62881-047, A62881-048, A62881-049, A62881-050, A62881-051
- Distribution
- US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Memphis, TN, United States