Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system

FDA recall source

Recall number
Z-0141-2027
Event id
99774
Status
Ongoing
FDA classification
Class II
Recalling firm
Cadwell Industries Inc
Report date
Oct 21, 2026
Posted
Oct 9, 2026
Initiated
Aug 18, 2026
Center classification date
Oct 9, 2026
Product code
GWF
Device type
Stimulator, Electrical, Evoked Response
Reason
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
Root cause
Under Investigation by firm
Quantity
23
Code info
Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Lot Code: These devices are identified by the serial numbers listed below: 19029607AA0326007 19029607AA0226013 19029607AA0426028 19029607AA0226010 19029607AA0226007 19029607AA0326026 19029607AA0226015 19029607AA0326020 19029607AA0326018 19029607AA0326021 19029607AA0326022 19029607AA0226009 19029607AA0526022 19029607AA0326017 19029607AA0326027 19029607AA0326016 19029607AA0226016 19029607AA0326028 19029607AA0326024 19029607AA0326025 19029607AA0326023 19029607AA0326031 19029607AA0326006
Distribution
Worldwide - US Nationwide distribution in the state of TN and the countries of Spain, Vietnam, Malaysia, Saudi Arabia, Thailand, Israel, Mexico, Turkey, Oman, Colombia, Brazil, Belgium, Italy, South Africa, and Switzerland.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K162199
Location
Kennewick, WA, United States