Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;

FDA recall source

Recall number
Z-0143-2027
Event id
99829
Status
Ongoing
FDA classification
Class II
Recalling firm
Boston Scientific Corporation
Report date
Oct 21, 2026
Posted
Oct 9, 2026
Initiated
Sep 10, 2026
Center classification date
Oct 9, 2026
Product code
FGE
Device type
Stents, Drains And Dilators For The Biliary Ducts
Reason
Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.
Root cause
Under Investigation by firm
Quantity
403
Code info
Material Number (UPN): M00535030; UDI-DI: 08714729856191; Lot Number: 38904309, 39017263, 39365322;
Distribution
US Nationwide distribution in the states of AL, CA, FL, GA, IL, IN, KY, MD, NC, NH, NM, NY, RI, TX, WA, WI, WV and the countries in TBD.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K223592
Location
Marlborough, MA, United States