Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm)

FDA recall source

Recall number
Z-0146-2027
Event id
99868
Status
Ongoing
FDA classification
Class II
Recalling firm
SI-BONE, Inc.
Report date
Oct 21, 2026
Posted
Oct 9, 2026
Initiated
Aug 21, 2026
Center classification date
Oct 9, 2026
Product code
OUR
Device type
Sacroiliac Joint Fixation
Reason
The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.
Root cause
Packaging
Quantity
100 units
Code info
10055T-UDI:00810055520077/Lot:9142871 10060T-UDI:00810055520084/Lot:9137761 10065T-UDI:00810055520091/Lot:9139213 10080T-UDI:00810055520121/Lot:9137781 11580T-UDI:00810055520367/Lot:9142911
Distribution
US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K203247
Location
Santa Clara, CA, United States