Generic Name: FilmArray¿ Pneumonia Panel Generic Name: FilmArray¿ Pneumonia Panel Model/Catalog Number: RFIT-ASY-0144 (30 pack), RFIT-ASY-0145 (6 pack), FLM1-ASY-0150 (Verification Kits). Verification kit lots (FLM1-ASY-0150) are a sales configuration that consume kits from panel lots (RFIT-ASY-0144). Software Version: N/A Product Description: Multiple-type respiratory pathogen nucleic acid IVD, kit, nucleic acid technique (NAT) Component: N/A

FDA recall source

Recall number
Z-0148-2027
Event id
99592
Status
Ongoing
FDA classification
Class II
Recalling firm
BioFire Diagnostics, LLC
Report date
Oct 21, 2026
Posted
Oct 13, 2026
Initiated
Aug 12, 2026
Center classification date
Oct 13, 2026
Product code
QDP
Device type
Respiratory Panel
Reason
Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results
Root cause
Under Investigation by firm
Quantity
4,892
Code info
Lot Code: UDI nos 00815381020178 (30 pack), 00815381020185 (6 pack); Model Nos RFIT-PRT-0144 (30 pack), RFIT-ASY-0145 (6 pack), FLM1-ASY-0150 (Verification Kit); Lot Numbers 0638726, 0676626, 0676726, 0716526, 0727826, 0767926, 0804526, 0804626, 0860926, 0882826, 0882926, 0883026, 0914626, 0914726, 0914826, 0914926, 0969126, 0969726, 0978026, 0978126, 0978226, 0978326, 0997626, 0997826, VK0997826, 0999226, 1020826, 1020926, 1105426, 1119126, 1119226, 1119326, 1119426, 1151826, 1152326, 1152426, 1179726, 1219026, 1267526, 1286226, 1286326, 1286426, 1346126, 1402726, 1402826, 1402926, 1419126, 1419226, 1437526, 1478926, 1560326, 1560426, 1560526, 1630726, 1656826, 1683726, 1702326, 1708826
Distribution
Worldwide - U.S. Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of United Arab Emirates, Netherland Antilles, Argentina, Austria, Bosnia and Herzegovina, Belgium, Bulgaria, Bolivia, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Czech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Spain, Finland, France, United Kingdom, Greece, Guatemala, Hong Kong, Honduras, Croatia, Hungary, Indonesia, Israel, India, Italy, Japan, Korea, Lebanon, Sri Lanka, Lithuania, Mexico, Malaysia, Nicaragua, Panama, Peru, Philippines, Poland, Portugal, Paraguay, Romania, Serbia, Singapore, El Salvador, Thailand, Taiwan, Venezuela, Vietnam.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K243222
Location
Salt Lake City, UT, United States