Device recall
Philips Lumify Diagnostic Ultrasound System
- Recall number
- Z-0152-2027
- Event id
- 99878
- Status
- Ongoing
- FDA classification
- Class I
- Recalling firm
- Philips Ultrasound, LLC
- Report date
- Oct 28, 2026
- Posted
- Oct 16, 2026
- Initiated
- Sep 24, 2026
- Center classification date
- Oct 16, 2026
- Product code
- IYN
- Device type
- System, Imaging, Pulsed Doppler, Ultrasonic
- Reason
- Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
- Root cause
- Device Design
- Quantity
- 13,206
- Code info
- UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Bothell, WA, United States