Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H

FDA recall source

Recall number
Z-0153-2027
Event id
99754
Status
Ongoing
FDA classification
Class II
Recalling firm
Ethicon, LLC
Report date
Oct 21, 2026
Posted
Oct 14, 2026
Initiated
Sep 8, 2026
Center classification date
Oct 14, 2026
Product code
GAQ
Device type
Suture, Nonabsorbable, Steel, Monofilament And Multifilament, Sterile
Reason
Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
Root cause
Process control
Quantity
1764
Code info
UDI-DI:Box/Unit: 30705031047369/10705031047365. Lot: RLBDPJ. Expiration: SEPTEMBER 30, 2026
Distribution
Worldwide - US Nationwide distribution in the states of SC, PA, TX, CA, ID, LA, MO, AL, NC, UT, IL, MA, OH, AZ, TN, OK, NE, ME, IN, WI, FL, CO, WY, NH, NY, NC, MN and the country of CA.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K946173
Location
San Lorenzo, PR, United States