Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm

FDA recall source

Recall number
Z-0157-2027
Event id
99846
Status
Ongoing
FDA classification
Class II
Recalling firm
In2bones USA, LLC
Report date
Oct 21, 2026
Posted
Oct 14, 2026
Initiated
Sep 9, 2026
Center classification date
Oct 14, 2026
Product code
HWC
Device type
Screw, Fixation, Bone
Reason
Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
Root cause
Error in labeling
Quantity
29 units
Code info
UDI: 00817906023992/Lot: GM008380; GM007868
Distribution
US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K170518
Location
Cordova, TN, United States