Device recall

LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.

FDA recall source

Recall number
Z-0162-2027
Event id
99776
Status
Ongoing
FDA classification
Class III
Recalling firm
Magellan Diagnostics, Inc.
Report date
Oct 21, 2026
Posted
Oct 15, 2026
Initiated
Sep 4, 2026
Center classification date
Oct 15, 2026
Product code
DOF
Device type
Lead, Atomic Absorption
Reason
In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
Root cause
Labeling Change Control
Quantity
321 units
Code info
Model/Catalog Number: 70-6762. UD-DI: 00850355006000. Lot Number. 2530M-01.
Distribution
US Nationwide distribution in the states of CA, IN, PA, TX, NE, IL, MA, MO, KY.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K052549
Location
North Billerica, MA, United States