Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112

FDA recall source

Recall number
Z-0163-2027
Event id
99875
Status
Ongoing
FDA classification
Class II
Recalling firm
Cook Incorporated
Report date
Oct 21, 2026
Posted
Oct 15, 2026
Initiated
Sep 16, 2026
Center classification date
Oct 15, 2026
Product code
FBK
Device type
Endoscopic Injection Needle, Gastroenterology-Urology
Reason
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
Root cause
Under Investigation by firm
Quantity
97 units
Code info
UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X
Distribution
US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K171602
Location
Bloomington, IN, United States