Device recall

Medtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly

FDA recall source

Recall number
Z-0195-2027
Event id
99895
Status
Ongoing
FDA classification
Class II
Recalling firm
Medtronic Navigation, Inc.
Report date
Oct 28, 2026
Posted
Oct 16, 2026
Initiated
Aug 31, 2026
Center classification date
Oct 16, 2026
Product code
HAW
Device type
Neurological Stereotaxic Instrument
Reason
Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Root cause
Under Investigation by firm
Quantity
83,594 units
Code info
UDI: 20805832070023/ Lot codes: 2503141 2503241 2503411 2503491 2504081 2504091 2504211 2504301 2504331 2505211 2505221 2505351 2505641 2505651 2505681 2506081 2506161 2506341 2506551 2506581 2509781 2509791 2509811 2510121 2510741 2510751 2511101 2511121 2511501 2511601
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K033621
Location