Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)

FDA recall source

Recall number
Z-0200-2027
Event id
99518
Status
Ongoing
FDA classification
Class II
Recalling firm
Greiner Bio-One North America, Inc.
Report date
Oct 28, 2026
Posted
Oct 21, 2026
Initiated
Jul 29, 2026
Center classification date
Oct 21, 2026
Product code
JKA
Device type
Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
Root cause
Nonconforming Material/Component
Quantity
860450
Code info
Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
Distribution
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K960857
Location
Monroe, NC, United States