Device recall

CareLink Clinic, REF: MMT-7350

FDA recall source

Recall number
Z-0203-2027
Event id
99765
Status
Ongoing
FDA classification
Class II
Recalling firm
Medtronic MiniMed, Inc.
Report date
Oct 28, 2026
Posted
Oct 21, 2026
Initiated
Sep 11, 2026
Center classification date
Oct 21, 2026
Product code
PHV
Device type
Continuous Glucose Monitor Retrospective Data Analysis Software
Reason
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
Root cause
Software design
Quantity
1
Code info
UDI-DI: 0763000B00008748K. Software Versions: 4.4A Software Available: US: Between June 4th, 2026, and June 8th, 2026.
Distribution
US Nationwide distribution in the state of KS.
Voluntary or mandated
Voluntary: Firm initiated
Location
Northridge, CA, United States