Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Techno-Path Multichem AE Unassayed Level 1 REF:AE621X

FDA recall source

Recall number
Z-0205-2027
Event id
99724
Status
Ongoing
FDA classification
Class II
Recalling firm
TECHNO-PATH MANUFACTURING LTD.
Report date
Oct 28, 2026
Posted
Oct 22, 2026
Initiated
Aug 14, 2026
Center classification date
Oct 22, 2026
Product code
OHQ
Device type
Multi-Analyte Controls Unassayed
Reason
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Root cause
Under Investigation by firm
Quantity
832 units
Code info
UDI: (01)05391523442800/Lot # 65705261
Distribution
US Nationwide distribution in the state of Massachusetts.
Voluntary or mandated
Voluntary: Firm initiated
Location
Ballina Tipperary, Ireland