Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use implant grade metal device, available in varied articulating surface coverages. The restor3d MTP implant is designed to replace the damaged articulating surface of the first metatarsal distal head. The device is intended for the treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock and integrity of the first metatarsal head. This implant can also be used in the following clinical conditions: hallux valgus or hallux limitus, hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The implant is a single use product intended to be press fit, with optional use of bone cement. The restor3d MTP Implant is comprised of a single laser powder bed fusion (LPBF) printed cobalt-chromium alloy (Co28Cr6Mo) part. The external surface of the implant that is intended to be in contact with the phalanx is mirror polished. Component: Not a component

FDA recall source

Recall number
Z-0264-2027
Event id
99940
Status
Ongoing
FDA classification
Class II
Recalling firm
restor3d Inc.
Report date
Oct 28, 2026
Posted
Oct 22, 2026
Initiated
Sep 4, 2026
Center classification date
Oct 22, 2026
Product code
KWD
Device type
Prosthesis, Toe, Hemi-, Phalangeal
Reason
Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
Root cause
Process change control
Quantity
16
Code info
Lot Code: Model: 1103-161620 UDI: 00840097505244 Lot: A02979
Distribution
U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K201393
Location
Durham, NC, United States