Device recall

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C

FDA recall source

Recall number
Z-0289-2027
Event id
99889
Status
Ongoing
FDA classification
Class II
Recalling firm
Canon Medical System, USA, INC.
Report date
Nov 4, 2026
Posted
Oct 27, 2026
Initiated
Oct 12, 2026
Center classification date
Oct 27, 2026
Product code
OWB
Device type
Interventional Fluoroscopic X-Ray System
Reason
Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
Root cause
Software design
Quantity
15 SYSTEMS
Code info
DI-04987670100192
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K210900
Location
Tustin, CA, United States