Device recall

EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated

FDA recall source

Recall number
Z-0301-2027
Event id
99931
Status
Ongoing
FDA classification
Class II
Recalling firm
BioMerieux SA
Report date
Nov 4, 2026
Posted
Oct 27, 2026
Initiated
Sep 11, 2026
Center classification date
Oct 27, 2026
Product code
JJH
Device type
Clinical Sample Concentrator
Reason
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
Root cause
Software change control
Quantity
33
Code info
UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351
Distribution
Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
Voluntary or mandated
Voluntary: Firm initiated
Location
Marcy L'Etoile, France