Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Precision Pointer Electromagnetic (8000-050-001)

FDA recall source

Recall number
Z-0314-2027
Event id
99855
Status
Ongoing
FDA classification
Class II
Recalling firm
Stryker Leibinger GmbH & Co. KG
Report date
Nov 4, 2026
Posted
Oct 29, 2026
Initiated
Sep 24, 2026
Center classification date
Oct 29, 2026
Product code
HAW
Device type
Neurological Stereotaxic Instrument
Reason
Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
Root cause
Process control
Quantity
513
Code info
Lot 7000413361 UDI-DI 07613327413205
Distribution
Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K212194
Location
Freiburg Im Breisgau, Germany