Device recall

Senographe Pristina 3D

FDA recall source

Recall number
Z-0328-2027
Event id
100070
Status
Open, Classified
FDA classification
Class III
Recalling firm
GE Medical Systems, LLC
Posted
Nov 4, 2026
Initiated
Aug 28, 2026
Product code
OTE
Device type
Digital Breast Tomosynthesis
Reason
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
Root cause
Radiation Control for Health and Safety Act
Code info
N/A
Distribution
USA and OUS.
Location
Waukesha, WI