Device recall
TruSystem 7000 U Operating Table (dV), Product Code 1723633
- Recall number
- Z-0332-2027
- Event id
- 99826
- Status
- Open, Classified
- FDA classification
- Class I
- Recalling firm
- Baxter Healthcare Corporation
- Posted
- Nov 3, 2026
- Initiated
- Sep 14, 2026
- Product code
- GDC
- Device type
- Table, Operating-Room, Electrical
- Reason
- The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
- Root cause
- Equipment maintenance
- Quantity
- 62 units
- Code info
- UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662
- Distribution
- US Nationwide distribution in the states of IL, MO, TX.
- Location
- Round Lake, IL