Device recall

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

FDA recall source

Recall number
Z-0348-2027
Event id
100047
Status
Open, Classified
FDA classification
Class II
Recalling firm
restor3d Inc.
Posted
Nov 3, 2026
Initiated
Oct 1, 2026
Product code
PHX
Device type
Shoulder Prosthesis, Reverse Configuration
Reason
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
Root cause
Device Design
Quantity
240 units
Code info
All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409
Distribution
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
Related 510(k)s
K243643
Location
Durham, NC