Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.

FDA recall source

Recall number
Z-0349-2027
Event id
100098
Status
Open, Classified
FDA classification
Class III
Recalling firm
Cochlear Limited
Posted
Nov 3, 2026
Initiated
Sep 30, 2026
Product code
MCM
Device type
Implant, Cochlear
Reason
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
Root cause
Under Investigation by firm
Quantity
4 device in the U.S.
Code info
Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
Distribution
U.S.: NJ, NY, and PA. O.U.S.: Not provided
Location
Macquarie University, Australia