Device recall
Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens
- Recall number
- Z-0355-2027
- Event id
- 100139
- Status
- Open, Classified
- FDA classification
- Class II
- Recalling firm
- Remel, Inc
- Posted
- Nov 3, 2026
- Initiated
- Oct 19, 2026
- Product code
- NGZ
- Device type
- Susceptibility Test Plate, Antifungal
- Reason
- Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
- Root cause
- Under Investigation by firm
- Quantity
- 33 US, 54 OUS
- Code info
- YY3462 lot 440400 UDI-DI 848838018817
- Distribution
- Worldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
- Related 510(k)s
- K090968
- Location
- Lenexa, KS