Device recall
LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303
- Recall number
- Z-0359-2027
- Event id
- 100036
- Status
- Open, Classified
- FDA classification
- Class II
- Recalling firm
- Luminex Corporation
- Posted
- Nov 5, 2026
- Initiated
- Oct 5, 2026
- Product code
- NSU
- Device type
- Instrumentation For Clinical Multiplex Test Systems
- Reason
- Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.
- Root cause
- Error in labeling
- Quantity
- 30,168 test cartridges (2,514 Kits)
- Code info
- All lots UDI-DI pending
- Distribution
- US distribution nationwide. No OUS distribution.
- Related 510(k)s
- K243013
- Location
- Northbrook, IL