Device recall

LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303

FDA recall source

Recall number
Z-0359-2027
Event id
100036
Status
Open, Classified
FDA classification
Class II
Recalling firm
Luminex Corporation
Posted
Nov 5, 2026
Initiated
Oct 5, 2026
Product code
NSU
Device type
Instrumentation For Clinical Multiplex Test Systems
Reason
Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.
Root cause
Error in labeling
Quantity
30,168 test cartridges (2,514 Kits)
Code info
All lots UDI-DI pending
Distribution
US distribution nationwide. No OUS distribution.
Related 510(k)s
K243013
Location
Northbrook, IL