Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics

FDA recall source

Recall number
Z-2927-2026
Event id
99450
Status
Ongoing
FDA classification
Class II
Recalling firm
Remel Europe Ltd.
Report date
Aug 19, 2026
Initiated
Jul 17, 2026
Center classification date
Aug 7, 2026
Reason
Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.
Quantity
26 units
Code info
UDI/DI 05056080501161, Lot Numbers: 6214791, exp. 18/03/2025; 6175327, exp. 04/12/2027; 6148809, exp. 30/09/2027; 6002010, exp. 31/07/2027.
Distribution
US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Location
Dartford, United Kingdom