Device recall

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

FDA recall source

Recall number
Z-2938-2026
Event id
99341
Status
Ongoing
FDA classification
Class II
Recalling firm
Erbe USA Inc
Report date
Aug 19, 2026
Initiated
Jun 26, 2026
Center classification date
Aug 12, 2026
Reason
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Quantity
1126 units
Code info
UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
Distribution
US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.
Voluntary or mandated
Voluntary: Firm initiated
Location
Marietta, GA, United States