Device recall
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
- Recall number
- Z-2959-2026
- Event id
- 99410
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Tobii Dynavox Llc
- Report date
- Aug 26, 2026
- Initiated
- Jul 8, 2026
- Center classification date
- Aug 14, 2026
- Reason
- Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
- Code info
- UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).
- Distribution
- US Nationwide distribution.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Pittsburgh, PA, United States