Device recall

TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

FDA recall source

Recall number
Z-2960-2026
Event id
99410
Status
Ongoing
FDA classification
Class II
Recalling firm
Tobii Dynavox Llc
Report date
Aug 26, 2026
Initiated
Jul 8, 2026
Center classification date
Aug 14, 2026
Reason
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Code info
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
Distribution
US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Location
Pittsburgh, PA, United States