Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001

FDA recall source

Recall number
Z-2967-2026
Event id
99476
Status
Ongoing
FDA classification
Class II
Recalling firm
Paragonix Technologies Inc.
Report date
Aug 26, 2026
Initiated
Jul 20, 2026
Center classification date
Aug 14, 2026
Reason
Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Quantity
1445 systems
Code info
All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2
Distribution
US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR.
Voluntary or mandated
Voluntary: Firm initiated
Location
Braintree, MA, United States