Device recall
Custom Made Device Implant 6123 Navicular Cage Nominal Right Rev A, Model/Catalog Number 6123N R; foot implant
- Recall number
- Z-2972-2026
- Event id
- 99504
- Status
- Ongoing
- FDA classification
- Class III
- Recalling firm
- Paragon 28, Inc.
- Report date
- Aug 26, 2026
- Initiated
- Jul 10, 2026
- Center classification date
- Aug 14, 2026
- Reason
- Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
- Quantity
- 1 unit
- Code info
- Lot Number AOCIA6123
- Distribution
- US Nationwide distribution i the state of MT.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Englewood, CO, United States