Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

FDA recall source

Recall number
Z-2973-2026
Event id
99436
Status
Ongoing
FDA classification
Class II
Recalling firm
Beckman Coulter Inc.
Report date
Aug 26, 2026
Posted
Aug 17, 2026
Initiated
Jul 15, 2026
Center classification date
Aug 17, 2026
Product code
JJE
Device type
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Root cause
Software design (manufacturing process)
Code info
1) Part Number C63522, UDI/DI 14987666545089; 2) Part Number C63521, UDI/DI 14987666545072; 3) Part Number C13252, UDI/DI 15099590742331.
Distribution
Worldwide distribution - US Nationwide.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K242190
Location
Brea, CA, United States