Device recall

Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL

FDA recall source

Recall number
Z-2976-2026
Event id
99446
Status
Ongoing
FDA classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Report date
Aug 26, 2026
Posted
Aug 17, 2026
Initiated
Jul 13, 2026
Center classification date
Aug 17, 2026
Product code
LBZ
Device type
Enzyme Immunoassay, Quinidine
Reason
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
Root cause
Unknown/Undetermined by firm
Quantity
55 units
Code info
Lots: 2597, 2627, 2709/UDI: 00842768023293
Distribution
Worldwide - US Nationwide distribution in the state of CA and the country of Italy.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K011605
Location
East Walpole, MA, United States