Device recall
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
- Recall number
- Z-2980-2026
- Event id
- 99374
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Medtronic MiniMed, Inc.
- Report date
- Aug 26, 2026
- Posted
- Aug 18, 2026
- Initiated
- Jul 13, 2026
- Center classification date
- Aug 18, 2026
- Product code
- QFG
- Device type
- Alternate Controller Enabled Insulin Infusion Pump
- Reason
- Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
- Root cause
- Process design
- Quantity
- 16
- Code info
- Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H
- Distribution
- International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
- Voluntary or mandated
- Voluntary: Firm initiated
- Related 510(k)s
- K251032
- Location
- Northridge, CA, United States