Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;

FDA recall source

Recall number
Z-2981-2026
Event id
99374
Status
Ongoing
FDA classification
Class II
Recalling firm
Medtronic MiniMed, Inc.
Report date
Aug 26, 2026
Posted
Aug 18, 2026
Initiated
Jul 13, 2026
Center classification date
Aug 18, 2026
Product code
OZP
Device type
Automated Insulin Dosing Device System, Single Hormonal Control
Reason
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Root cause
Process design
Quantity
7
Code info
Model/UDI-DI Serial Numbers: MMT-1781/00763000404536/NG2665223H, NG2748200H, NG2748664H, NG2755237H, NG2716176H, NG3547582H. MMT-1782/00763000404543/NG2768190H
Distribution
International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
Voluntary or mandated
Voluntary: Firm initiated
Location
Northridge, CA, United States