Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW

FDA recall source

Recall number
Z-2982-2026
Event id
99374
Status
Ongoing
FDA classification
Class II
Recalling firm
Medtronic MiniMed, Inc.
Report date
Aug 26, 2026
Posted
Aug 18, 2026
Initiated
Jul 13, 2026
Center classification date
Aug 18, 2026
Product code
QFG
Device type
Alternate Controller Enabled Insulin Infusion Pump
Reason
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Root cause
Process design
Quantity
257
Code info
Model/UDI-DI Serial Numbers: MMT-1885CLW/00763000442187, 00763000974374/NG3150737H, NG3155194H, NG3151561H, NG3154693H, NG3151716H, NG3153547H, NG3153637H, NG3155407H, NG3152666H, NG3155788H, NG3081975H, NG2901117H, NG3080842H, NG3081463H, NG3153416H, NG3081894H, NG3153781H, NG3155944H, NG3081982H, NG3081525H, NG3155909H, NG3090722H, NG3154897H, NG3088436H, NG3088708H, NG3090073H, NG3155908H, NG3150720H, NG3092957H, NG3128715H, NG3128784H, NG3152774H, NG3153929H, NG3514566H, NG3150780H, NG3151290H, NG3151508H, NG3507161H, NG3508693H, NG3081976H, NG3151460H, NG3152051H, NG3152085H, NG3152344H, NG3153213H, NG3152772H, NG3155586H, NG3150786H, NG3153214H, NG3153385H, NG3151840H, NG3082019H, NG3153405H, NG3153788H, NG3153908H, NG3154527H, NG3151598H, NG2854971H, NG3155196H, NG3155202H, NG3155341H, NG3091491H, NG3155557H, NG3081498H NG3155688H, NG3155020H, NG3508565H, NG3155419H, NG3507536H, NG3091563H, NG3082036H, NG3154689H, NG3154995H, NG3508504H, NG3514345H, NG3509850H, NG3527505H, NG3514547H, NG3527469H, NG3504333H, NG3504598H, NG3505071H, NG3505220H, NG3505273H, NG3514617H, NG3506727H, NG3085044H, NG3507542H, NG3507547H, NG3514448H, NG3508303H, NG3508350H, NG3508371H, NG3514612H, NG3508511H, NG3508617H, NG3508698H, NG3508707H, NG3508757H, NG3508799H, NG3508903H, NG3509138H, NG3509295H, NG3509299H, NG3509452H, NG3509685H, NG3509689H, NG3509836H, NG3509914H, NG3509925H, NG3514375H, NG3514441H, NG3505246H, NG3505166H, NG3505283H, NG3151570H, NG3154948H, NG3514667H, NG3527165H, NG3527243H, NG3527298H, NG3527468H, NG3527480H, NG3527491H, NG3527580H, NG3527529H, NG3527543H, NG3527571H, NG3527574H, NG3092722H, NG3527581H, NG3527611H, NG3527637H, NG3527642H, NG3527672H, NG3527710H, NG3527737H, NG3527760H, NG3527865H, NG3527918H, NG3527961H, NG3527989H, NG3528017H, NG3528023H, NG3528036H, NG3528061H, NG3528120H, NG3528147H, NG3528156H, NG3528158H, NG3528180H, NG3528218H, NG3528226H, NG3528378H, NG3539964H, NG3539969H, NG3539970H, NG3539999H, NG3540001H, NG3540013H, NG3540029H, NG3540046H, NG3540053H, NG3540084H, NG3540191H, NG3540270H, NG3540310H, NG3540320H, NG3540581H, NG3540604H, NG3540628H, NG3540629H MMT-1886CLW/00763000442194, 00763000974381/NG2966956H, NG2970938H, NG2986362H, NG3013390H, NG3057910H, NG3167493H, NG3466128H, NG3098147H, NG3136398H, NG3137113H, NG3139432H, NG3143920H, NG3144721H, NG3145486H, NG3145488H, NG3471528H, NG3167609H, NG3147120H, NG3147035H, NG3430059H, NG3438451H, NG3464758H, NG3163796H, NG3166778H, NG3409587H, NG3167450H, NG3167623H, NG3167967H, NG3169694H, NG3170253H, NG3321081H, NG3409583H, NG3409611H, NG3409656H, NG3409680H, NG3409697H, NG3409716H, NG3409807H, NG3409825H, NG3409831H, NG3409854H, NG3409925H, NG3409959H, NG3410005H, NG3410053H, NG3410092H, NG3410111H, NG3410166H, NG3410189H, NG3410227H, NG3410233H, NG3410243H, NG3410279H, NG3410304H, NG3410389H, NG3410398H, NG3429103H, NG3429377H, NG3429485H, NG3429881H, NG3430107H, NG3430448H, NG3437945H, NG3437957H, NG3438394H, NG3470788H, NG3466185H, NG3466619H, NG3466620H, NG3467368H, NG3467586H, NG3468759H, NG3469430H, NG3469907H, NG3470728H, NG3168104H, NG3472157H, NG3169669H, NG3170079H, NG3430111H, NG3429099H, NG3437621H, NG3167969H, NG3465813H, NG3167397H
Distribution
International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K251032
Location
Northridge, CA, United States