Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

FDA recall source

Recall number
Z-2983-2026
Event id
99371
Status
Ongoing
FDA classification
Class II
Recalling firm
Fort Defiance Industries, LLC
Report date
Aug 26, 2026
Posted
Aug 18, 2026
Initiated
Jul 8, 2026
Center classification date
Aug 18, 2026
Product code
FLE
Device type
Sterilizer, Steam
Reason
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Root cause
Component design/selection
Code info
Lot Code: All serial numbers/ GTIN: (01) 00862992000327
Distribution
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K141009
Location
Loudon, TN, United States