Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier.

FDA recall source

Recall number
Z-2991-2026
Event id
99403
Status
Ongoing
FDA classification
Class II
Recalling firm
Medline Industries, LP
Report date
Sep 2, 2026
Posted
Aug 21, 2026
Initiated
Jul 13, 2026
Center classification date
Aug 21, 2026
Product code
BTT
Device type
Humidifier, Respiratory Gas, (Direct Patient Interface)
Reason
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Root cause
Under Investigation by firm
Quantity
20 units
Code info
All lot codes
Distribution
Worldwide distribution - US Nationwide and the country of Panama.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K031383
Location
Northfield, IL, United States