Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

FDA recall source

Recall number
Z-3013-2026
Event id
99404
Status
Ongoing
FDA classification
Class II
Recalling firm
Midmark Corporation
Report date
Sep 2, 2026
Posted
Aug 25, 2026
Initiated
Jul 20, 2026
Center classification date
Aug 25, 2026
Product code
DRG
Device type
Transmitters And Receivers, Physiological Signal, Radiofrequency
Reason
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
Root cause
Software design
Quantity
3624 units
Code info
Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark Digital ECG. The Midmark Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 7817519 7817645, L1130587 V2967564 UDI-DI: 00841709121319, 08056732652491
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K220526
Location
Versailles, OH, United States