Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

FDA recall source

Recall number
Z-3014-2026
Event id
99222
Status
Ongoing
FDA classification
Class II
Recalling firm
Canon Medical System, USA, INC.
Report date
Sep 2, 2026
Posted
Aug 25, 2026
Initiated
Aug 4, 2026
Center classification date
Aug 25, 2026
Product code
OWB
Device type
Interventional Fluoroscopic X-Ray System
Reason
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Root cause
Process control
Quantity
10 systems
Code info
Model Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: Pending SID Number: 30059180 Serial Number: STA2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30095017 Serial Number: SZA2432003 SID Number: 30097052 Serial Number: SZA2422002
Distribution
U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K182415K203551K210900
Location
Tustin, CA, United States