Device recall

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

FDA recall source

Recall number
Z-3033-2026
Event id
99596
Status
Ongoing
FDA classification
Class II
Recalling firm
IMRIS Imaging Inc
Report date
Sep 2, 2026
Posted
Aug 26, 2026
Initiated
Aug 6, 2026
Center classification date
Aug 26, 2026
Product code
FQO
Device type
Table, Operating-Room, Ac-Powered
Reason
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Root cause
Nonconforming Material/Component
Quantity
6 units
Code info
113821-600 - Serial Numbers: 10003754, 10006997; 114148-000 - UDI-DI: 00857534006349, Serial Numbers: 10005189, 10005190, 10004463, 10004997
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Voluntary or mandated
Voluntary: Firm initiated
Location
Chaska, MN, United States