Device recall

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

FDA recall source

Recall number
Z-3034-2026
Event id
99596
Status
Ongoing
FDA classification
Class II
Recalling firm
IMRIS Imaging Inc
Report date
Sep 2, 2026
Posted
Aug 26, 2026
Initiated
Aug 6, 2026
Center classification date
Aug 26, 2026
Product code
KXJ
Device type
Table, Radiologic
Reason
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Root cause
Nonconforming Material/Component
Quantity
13 units
Code info
114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Voluntary or mandated
Voluntary: Firm initiated
Location
Chaska, MN, United States