Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209

FDA recall source

Recall number
Z-3036-2026
Event id
99536
Status
Ongoing
FDA classification
Class II
Recalling firm
GE Medical Systems Information Technologies Inc
Report date
Sep 2, 2026
Posted
Aug 27, 2026
Initiated
Jul 20, 2026
Center classification date
Aug 27, 2026
Product code
MWI
Device type
Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Reason
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
Root cause
Device Design
Quantity
23 units
Code info
GTIN 00195278679284, Serial Numbers: SWT25250001TA, SWT25360001TA, SWT25360002TA, SWT24090006TA, SWT25100001TA, SWT25120002TA, SWT24140007TA, SWT24150008TA, SWT25100004TA, SWT24050001TA, SWT25110003TA, SWT24180011TA, SWT25120003TA, SWT25230001TA, SWT25230002TA, SWT25110009TA, SWT25110010TA, SWT25230003TA, SWT25230005TA, SWT25230006TA, SWT25230004TA, SWT25260001TA, SWT31240012TA.
Distribution
US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K234130
Location
Waukesha, WI, United States